TB Research

A Novel Medical Device for Airway Clearance

Nir Helper, Moshe Ashkenazi, G. V. Sokol, Adi Dagan, Ori Efrati

Journal of Clinical Medicine · 2025-01

Abstract

Background: Airway clearance techniques are a key element in the daily treatment of people with bronchiectasis. There are several methods and devices to assist in effective airway clearance. We investigated LibAirty, a novel medical device, and compared it with the common practice performed today. Methods: Twenty adults enrolled, and each one had three different treatments in a randomized order: a human respiratory physiotherapist, a High-Frequency Chest Wall Oscillator, and LibAirty with BiPAP. The outcome parameters were mucus weight and a questionnaire. Further studies were performed to investigate LibAirty with hypertonic saline (HS) inhalation and using the device as a standalone. Results: No adverse events were recorded. The sputum amount expectorated in all arms using LibAirty was 14.4 ± 11.1 g with BIPAP, 16.4 ± 7 g with HS, and 11.3 ± 4.1 g for the standalone treatment. For HFCWO, 4.45 ± 3.28 g was obtained, and for CPT, 15.9 ± 11.1 g was obtained. The amount obtained by LibAirty (all arms) was significantly higher than HFCWO. Conclusions: All arms of LibAirty were superior to HFCWO and similar to the human physiotherapist. Further studies should be performed to investigate the long-term effects of LibAirty.

MeSH terms

  • Medicine
  • Hypertonic saline
  • Airway
  • Sputum
  • Anesthesia
  • Adverse effect
  • Saline
  • Inhalation