Evaluation of MetaSystems Automated Fluorescent Microscopy System for the Machine-Assisted Detection of Acid-Fast Bacilli in Clinical Samples
Gianna Tomasello, Farnaz Foroughi, Danielle Padron, Ángel Moreno, Niaz Banaei
Journal of Clinical Microbiology · 2022-09
Abstract
= 11) sources were evaluated with the MetaSystems Mycobacteria Scanner running the NEON Metafer AFB Module. The sensitivity and specificity of the MetaSystems platform was measured as a standalone diagnostic and as an assistant to technologists to review positive images. Culture results were used as the reference method. The MetaSystems platform failed to scan 57 (11.5%) slides. The MetaSystems platform used as a standalone had a sensitivity of 97.0% (129/133; 95% CI 92.5 to 99.2) and specificity of 12.7% (46/363; 95% CI 9.4 to 16.5). When positive scans were used to assist technologists, the MetaSystems platform had a sensitivity of 70.7% (94/133; 95% CI 62.2 to 78.3) and specificity of 89.0% (323/363; 95% CI 85.3 to 92.0). The manual microscopy method had a sensitivity of 79.7% (106/133; 95% CI 71.9 to 86.2) and specificity of 98.6% (358/363; 95% CI 96.8 to 99.6). The sensitivity of the MetaSystems platform was not impacted by smear grade or mycobacterial species. The majority (70.3%) of false positive smears had ≥2+ smear results with the MetaSystems platform. Further performance improvements are needed before the MetaSystems' automated fluorescent AFB slide reader can be used to assist microscopist in the clinical laboratory.