Comparative study of a modified double-lumen tube ventilation control connector and traditional connector in clinical use: a randomised-controlled trial

Chang Liu, Yuanyu Zhao, You Li, Huiwen Guan, Junjie Feng, Shengquan Cheng, Xin Wang, Yue Wang, et al. (9 authors)

BMC Anesthesiology · 2022-09

Abstract

BACKGROUND: A Y-shaped rotatable connector (YRC) for double-lumen tubes (DLT) is invented and compared with the traditional connector (Y-shaped connector, YC). METHODS: Sixty patients with ASA grade I-III, aged ≥ 18 years, who needed to insert a DLT for thoracic surgery were recruited and assigned into the YRC group (n = 30) and the YC group (n = 30) randomly. The primary endpoints included the inhaled air concentration (Fi) and the exhaled air concentration (Et) of sevoflurane before and after the switch between two-lung ventilation and one-lung ventilation at different times, positioning time, and switching time. The secondary endpoints were the internal gas volume of the two connectors, airway pressure, and the sputum suction time. RESULTS: The Et and Fi of the YRC group and the YC group were significantly different (all p < 0.05) at 5s, 10s, and 30s after the patient switched from two-lung ventilation to one-lung ventilation. The positioning time of the YRC group was less than YC group (89.75 ± 14.28 s vs 107.80 ± 14.96 s, p < 0.05), as well as the switching time (3.60 ± 1.20 s vs 9.05 ± 2.53 s, p < 0.05) and the internal gas volume (17.20 ml vs 24.12 ml). There was no difference in airway pressure and the sputum suction time in two groups. CONCLUSION: Compared with YC, YRC was beneficial for maintaining depth of anesthesia, improves efficiency for the switch between one-lung and two-lung ventilation, and shortens the tube positioning time.

MeSH terms

  • Medicine
  • Anesthesia
  • Ventilation (architecture)
  • Peak inspiratory pressure
  • Tidal volume
  • Airway
  • Lung
  • Lumen (anatomy)
  • Sputum
  • Suction
  • Mechanical ventilation
  • Surgery