Development and evaluation of a multiplex quantitative polymerase chain reaction assay for detecting bacteria associated with lower respiratory tract infection
Xianxia Zhuo, Jiankang Zhao, Lei Wang, Bin Sun, Lanhua Sun, Chunlei Wang, Binbin Li, Yanyan Fan, et al. (10 authors)
International Journal of Infectious Diseases · 2022-05
Abstract
OBJECTIVES: This study aimed to establish a multiplex quantitative polymerase chain reaction (MQ-PCR) assay for 12 bacterial pathogens found in lower respiratory tract infection (LRTI) and to evaluate its performance in a cohort of 211 patients with LRTI. METHODS: The study was divided into two stages: a pilot study to establish the methodology and a clinical validation study to evaluate its performance. In the pilot study, we established the MQ-PCR and analyzed its performance regarding limits of detection, reproducibility, specificity, and efficiency. In the clinical validation study, we obtained 211 sputum and/or bronchoalveolar lavage fluid (BALF) samples and detected pathogens by MQ-PCR. The MQ-PCR time was 3 h from sample collection to complete pathogen detection. RESULTS: copies/ml in BALF were applied, the sensitivity, specificity, and positive and negative predictive values of the MQ-PCR were 77% (95% confidence interval [CI] 67-88%), 94% (95% CI 93-95%), 25% (95% CI 19-31%), and 99% (95% CI 99-100%), respectively. CONCLUSIONS: This study demonstrates that the new MQ-PCR assay is time-saving, more effective and sensitive, and brings us closer to mainstream adoption of quantitative molecular detection of bacteria.
MeSH terms
- Sputum
- Multiplex polymerase chain reaction
- Polymerase chain reaction
- Bronchoalveolar lavage
- Multiplex
- Respiratory tract infections
- Lower respiratory tract infection
- Confidence interval
- Real-time polymerase chain reaction
- Detection limit
- Bacteria
- Biology
- Medicine
- Microbiology