Multiplex Detection of Three Select Agents Directly from Blood by Use of the GeneXpert System
Padmapriya P. Banada, Srinidhi Deshpande, Sukalyani Banik, Darshini Shah, Ranie Koshy, Bhavana Patel, Robert Kwiatkowski, David H. Persing, et al. (9 authors)
Journal of Clinical Microbiology · 2019-03
Abstract
The sensitivity was 100% at the LOD for all three select agent bacteria in spiked patient blood samples. The assay specificity was 100% when it was tested against both nontarget pathogens and clinical patient blood samples. The total assay time was approximately 100 min. This automated assay, which is suitable for use at the point of care, identifies three select agents directly in blood without the need for enrichment with a high sensitivity within 100 min. This assay may enable rapid detection and treatment of patients infected with the target organisms in the event of a bioterrorism attack.